Method Development and Validation for Elemental Impurities in Deflazacort Tablet Dosage Form by Inductively Coupled Plasma Mass Spectrometry

 

Vijay Patel*, Nishu, Jaswandi Mehetre, Tushar Mehta, Kalpesh Upadhyay

School of Pharmacy, ITMBU University, Halol Highway, Vadodara 391510, Gujarat, India.

*Corresponding Author E-mail: Vijay293patel@gmail.com

 

ABSTRACT:

Trace-level elemental impurities in pharmaceuticals pose serious toxicity risks, requiring strict controls to safeguard patients and meet regulatory demands. From January 2018, USP <232> and ICH Q3D supplanted the legacy heavy metals test, specifying sensitive, selective techniques such as ICP-OES and ICP-MS for elemental impurities.1 Elemental impurity controls now form a required element of all regulatory filings. This study developed and validated a sensitive, precise ICP-MS method to quantify elemental impurities in Deflazacort tablets. Microwave digestion with dilute nitric acid prepared samples, extracting tablet matrix impurities effectively while cutting reagent needs and boosting lab safety. The validated method simultaneously quantified key USP <232> elemental impurities for oral products, Class 1(As, Cd, Hg, Pb) and Class 2A(Co, V, Ni) against their PDE limits.  Method validation followed USP <233> guidelines, covering specificity, linearity, accuracy, precision, LOD, LOQ, and system suitability. The method successfully analyzed elemental impurities in Deflazacort tablets and oral suspension samples. Validation results confirmed regulatory compliance and backed several market authorization filings. Despite ICP-MS's common use in testing pharmaceutical elemental impurities, no prior validated method targets Deflazacort marketed formulations, as per literature review. The current work fills a key gap with its innovative approach, practical utility and clear research merit.

 

KEYWORDS: Inductively coupled plasma mass spectrometry (ICP-MS), Active pharmaceutical ingredients (APIs), Elemental Impurities (EI), Deflazacort (DF), International Council for Harmonisation (ICH), United States Pharmacopeia (USP), Permissible Daily Exposure (PDE), Maximum daily dose (MDD), Kinetic energy discrimination (KED), Internal standard (IS), Count-per-second (CPS).

 

 


 

INTRODUCTION:

Regulatory agencies now prioritize elemental impurities for their toxicity risks at trace amounts. Pharmaceutical quality controls must meet strict standards to confirm product safety for patients1. These impurities can enter drug products from a variety of sources, including raw materials, active pharmaceutical ingredients (APIs), excipients, catalysts employed during synthesis, manufacturing equipment, and packaging components2.  Their presence, even at minute concentrations, may lead to undesirable pharmacological or toxicological effects, thereby posing significant risks to patient safety; hence, regulators like the FDA, MHRA, and TGA, along with pharmacopoeias, require precise and reliable detection of elemental impurities. USP sets permitted daily exposures (PDEs, in mg/day) for these impurities based on administration routes oral, parenteral, or inhalation fully aligning with ICH Q3D guidelines2,3. ICH Q3D sorts of elements into four classes by toxicity (PDE levels) and assesses their likelihood in drug products, as detailed in Table 14,5.

 

Table 1. Classification of elemental impurities, as per regulatory guidelines

Classification of Elemental Impurities According to USP <232> / ICH Q3D

Class

Description

Elements

Class 1

Human toxicants with route-independent toxicity and a high probability of occurrence

Arsenic (As),

Cadmium (Cd),

Mercury (Hg),

Lead (Pb)

Class 2A

Human toxicants with route-dependent toxicity and a relatively high probability of occurrence

Cobalt (Co),

Nickel (Ni),

Vanadium (V)

Class 2B

Human toxicants with route-dependent toxicity and a lower probability of occurrence

Silver (Ag), Gold (Au), Iridium (Ir), Osmium (Os), Palladium (Pd), Platinum (Pt), Rhodium (Rh), Ruthenium (Ru), Selenium (Se), Thallium (Tl)

Class 3

Elements with relatively low toxicity by the oral route, but which may require consideration for parenteral or inhalation exposure

Barium (Ba), Chromium (Cr), Copper (Cu), Lithium (Li), Molybdenum (Mo), Antimony (Sb), Tin (Sn)

Other Elements

Elements for which PDEs have not been established due to low inherent toxicity and/or regional regulatory differences

Aluminum (Al), Boron (B), Calcium (Ca), Iron (Fe), Potassium (K), Magnesium (Mg), Manganese (Mn), Sodium (Na), Tungsten (W), Zinc (Zn)

 

Prior to 2018, USP <231> governed elemental impurities in drug substances using a heavy metals limit test that compared precipitate color visually to a 10ppm lead standard yet it fell short in sensitivity, accuracy, and precision relative to USP <232> procedures1. In January 2018, authorities rolled out updated techniques to fix these flaws in heavy metals screening for impurities, moving to PDEs tailored per element instead of general limits. USP then overhauled <231>, implementing <232> for Elemental Impurities Limits and <233> for Elemental Impurities Procedures (per USP <232>). USP <233> recommends ICP-OES and ICP-MS as top choices, since they alone can quantify all 24 ICH Q3D elements in one run3,6. Various alternate analytical techniques, such as flame atomic absorption spectrometry (F-AAS) and graphite furnace atomic absorption spectrometry (GF-AAS), X-ray fluorescence spectrometry XRF instrumental neutron activation analysis (INAA), ICP-AES, and ICP-MS, are currently employed for the analysis of pharmaceuticals for detecting metals and metalloids7,8. These alternatives align with Procedures 1 and 2 outlined in the USP's elemental impurities chapter. Emerging plasma-based tools like microwave plasma-AES, ICP-TOFMS, HR-ICPMS, and dual ICP-MS systems are expanding the toolkit for such analyses. Among them, ICP-MS stands out as the preferred option, delivering far superior detection limits over ICP-OES due to its minimal background noise (USP <1730>)9.

 

In general, method development for ICP-MS presents significant challenges, particularly in the sample preparation of solid oral dosage forms that contain a wide variety of excipients. These excipients differ markedly in their physicochemical properties, especially solubility, and several of them are insoluble in aqueous solution. Consequently, achieving a clear and homogeneous sample solution becomes difficult and often necessitates the use and optimisation of microwave-assisted wet digestion or other digestion-based extraction techniques. Second major challenge involves attaining acceptable recoveries at trace levels from solid oral formulations and selecting an appropriate internal standard to ensure accurate quantification. These considerations are critical for meeting the analytical expectations outlined in USP <233> and USP <1730>10.

 

Deflazacort (DF), a synthetic glucocorticoid, sees broad clinical use for managing inflammatory and autoimmune disorders. As an oral solid dosage form administered chronically in certain patient populations, DF tablets require rigorous evaluation for elemental impurities in accordance with current regulatory expectations. Here, an analytical procedure for DF, different formulations like tablets and oral suspension, is developed and validated to overcome the above-mentioned challenges11. Microwave-assisted digestion with diluted nitric and hydrofluoric acids, combined with a matrix-matched approach, resolves low spike recovery issues and enables simultaneous quantification of all Class 1 and 2A elements in one run12,13. The analytical procedure was validated for specificity, LOD, LOQ, linearity, accuracy, precision, robustness, and system suitability, which were evaluated to demonstrate the method's reliability and regulatory compliance. This study delivers a compliant ICP-MS method for routine QC screening of elemental impurities in Deflazacort formulations like tablets. It supports regulatory checks and upholds drug safety for patient use14.

 

MATERIALS AND METHODS:

Materials:

Concentrated nitric acid (70%) and hydrofluoric acid (47–51%) of trace metal grade were used for sample preparation and digestion. Nitric acid was obtained from J.T. Baker, and hydrofluoric acid from VWR Chemicals. Diluted nitric acid and hydrofluoric acid solutions were prepared using 18 MΩ·cm deionised Milli-Q water and were utilised throughout the study. Single‑element standard solutions of lead (Pb), arsenic (As), cadmium (Cd), mercury (Hg), cobalt (Co), vanadium (V), nickel (Ni), gold (Au), indium (In), scandium (Sc), germanium (Ge), rhodium (Rh), bismuth (Bi), and titanium (Ti), each with a certified concentration of 1000µg/mL, were obtained from Inorganic Ventures. These standards were used in the preparation of calibration solutions and internal standard mixtures for ICP‑MS analysis.

 

Deflazacort tablet samples were provided by Amneal Pharmaceuticals Pvt. Ltd. All test materials were stored according to the conditions specified in their respective certificates of analysis. Solutions prepared from the test materials were maintained at room temperature unless otherwise noted. All reagents and chemicals employed in the study were of trace‑analysis grade or equivalent quality. Volumetric glassware used for sample and standard preparation consisted of polymethyl pentene or polypropylene to minimize the risk of elemental contamination. All sample preparation steps and ICP‑MS measurements were conducted at Amneal Pharmaceuticals Pvt. Ltd., in accordance with current Good Manufacturing Practices (cGMP)15.

 

Instrumentation:

Sample digestion occurred in a closed-vessel microwave system. Element quantification relied on a Thermo Scientific™ iCAP™ RQ ICP-MS coupled with a Teledyne CETAC ASX-560 autosampler. Analyses ran in KED mode using helium collision gas to minimize polyatomic interferences effectively9. Before each session, daily instrument optimization confirmed suitability. Automated tuning refined torch alignment, detector voltage, and ion optics, boosting sensitivity, resolution, and stability over the full mass range. The standard operating conditions and data acquisition parameters employed for ICP-MS measurements are summarised in Table 216,17. ICP-MS analysis was performed using optimized conditions with RF power of 1550W, plasma gas flow of 14L/min, nebulizer gas flow of 1.13L/min, dilution gas flow of 0.8 L/min, and helium collision gas flow of 4.5mL/min in KED mode. Quartz torch, nebulizer, and spray chamber with nickel sampler and skimmer cones were used. The lens voltage was maintained at 152V, with SEM detection, 0.001s dwell time, and 10 sweeps per reading, providing stable and sensitive detection of elemental impurities.

 

RESULTS:

Method Development:

Method development tackled key hurdles, especially poor recoveries for select elements tied to matrix effects18. Prior to method development, we determined µg/g acceptance limits for each elemental impurity. These values, derived from ICH Q3D/USP<233>, informed digestion and dilution adjustments. From ICH Q3D and USP <233>PDEs, paired with the drug's MDD, we calculated corresponding µg/g concentration limits. These derived limits guided optimization of sample weights and dilution factors. This approach yielded a sturdy digestion protocol for elemental impurity testing14.

 

Elements Limit (µg/g) =    PDE/MDD

PDE = PDE Value in µg/day per ICH Q3D

MDD = Maximum Daily Dose (g/day)

 

Initial open-vessel acid digestion trials using different combinations of HNO₃, HCl, and HF produced turbid solutions, indicating incomplete matrix decomposition and unsuitable sample preparation for ICP-MS analysis. To improve digestion efficiency, closed-vessel microwave digestion was evaluated under different acid and temperature conditions19. A clear and homogeneous solution was successfully obtained using 10mL HNO₃ and 0.1mL HF with microwave digestion at 1800 W and 200°C for 20minutes, demonstrating effective sample decomposition suitable for elemental impurity analysis20. Digestion cycles began with a 20minute ramp to reach target microwave vessel temperatures. ICP-MS multi-element analysis utilized KED mode to minimize polyatomic spectral overlaps.

 

Selection of Internal Standard:

ICP-MS demands precise internal standard choices based on key selection criteria to support accurate quantification. Ideally, the IS should have an atomic mass close to that of the analyte so that mass-dependent effects, such as matrix-induced signal variations or instrumental drift, affect both species similarly. It should also exhibit ionisation behaviour comparable to the analyte; although matching first ionisation potentials can be helpful, mass similarity is generally more critical for reliable results11. Additionally, the IS should be minimally affected by polyatomic interferences and respond consistently within the sample matrix to correct for matrix effects. Finally, the internal standard element must not be naturally present in the sample to avoid signal overlap and ensure precise quantification21. Appropriate internal standards were selected to ensure accurate correction of instrumental drift, matrix effects, and signal fluctuations during ICP-MS analysis. The selection was primarily based on mass proximity to the analyte, ionization behavior, and stability of signal response under KED-mode operating conditions. Yttrium was selected for vanadium, cobalt, and nickel due to its intermediate mass and consistent signal stability. Germanium was chosen for arsenic because of its close mass range and its ability to compensate for matrix-induced signal suppression. Indium was used for cadmium owing to its comparable ionization characteristics and stable response, while bismuth was selected for mercury and lead to provide reliable correction in the higher mass region. The selected internal standards demonstrated suitable compatibility with the analytes and contributed to accurate quantification throughout the method validation study. The internal standard concentrations were optimised to provide stable, sufficiently high count-per-second (CPS) responses without detector saturation. At the 100% standard level, the analyte concentrations (in ppb) and corresponding responses (in CPS) were as follows.


 

Table 2. 100% Standard Concentrations (ppb) and Corresponding ICP-MS Responses (CPS) for the Analysed Elements and their Internal Standards used for Calibration

Name of Element

100% Standard concentration (ppb)

100% Standard Response (CPS)

Internal standard

 

Internal Standard concentration (ppb)

Internal Standard Response (CPS)12

Vanadium (V)

37.040

979790

89Y

 

20

1277712

Cobalt (Co)

18.520

1263111

Nickel (Ni)

74.080

1315407

Arsenic (As)

5.560

15239

73Ge

300

518617

Cadmium (Cd)

1.852

29075

115In

6

699748

Mercury (Hg)

11.120

335451

209Bi

2

710739

 

Lead (Pb)

1.852

200

 


A comparatively higher CPS response was observed for certain internal standards during preliminary evaluation; therefore, their concentrations were adjusted to achieve stable and proportional signal correction.

 

Optimised Method:

Parameters for Microwave Digestion:

Samples underwent microwave-assisted digestion in a system set to dedicated digestion mode.  We used a MAXI rotor holding 44 vessels per run for high throughput. An in-line sensor tracked and held digestion temps steady. Safety lock activated above 80°C on vessels. Controlled params broke down samples fully for ICP-MS; details in Table 7. Microwave digestion parameters were optimized to achieve complete matrix decomposition with minimal analyte loss22,23. The finalized digestion program was performed at 1800 W and 180°C for 30 minutes in two stages, followed by 25 minutes of cooling. These conditions consistently produced clear and particle-free solutions suitable for sensitive ICP-MS analysis of elemental impurities. Based on Sequence of trail, the method was optimized as follow and instrument parameter are presented in table 3.


 

Table 3. Optimized ICP-MS Method Parameters for Quantification of Selected Analytes.

Analyte

Quantify

Internal Standard

Fit type

weighting

Forcing

Used for semiQuant

Vanadium (V)

Yes

Yttrium

Linear

None

Blank

Yes

Cobalt (Co)

Yes

Yttrium

Linear

None

Blank

Yes

Nickel (Ni)

Yes

Yttrium

Linear

None

Blank

Yes

Arsenic (As)

Yes

Germanium

Linear

None

Blank

Yes

Cadmium (Cd)

Yes

Indium

Linear

None

Blank

Yes

Mercury (Hg)

Yes

Bismuth

Linear

None

Blank

Yes

Lead (Pb)

Yes

Bismuth

Linear

None

Blank

Yes

 


Preparation of Solutions:

Diluent-1: A volume of 50mL of concentrated nitric acid was slowly added to approximately 500mL of purified water in a suitable beaker. Subsequently, 0.3mL of the gold standard solution (1000µg/mL) was introduced, and the mixture was diluted to a final volume of 1000mL with water. The solution was thoroughly mixed to ensure homogeneity.

 

Diluent-2: About 0.3mL of gold standard (1000µg/mL) was pipetted into approximately 500mL purified water within an appropriate beaker. Brought to 1000mL final volume using water and mixed thoroughly to obtain a homogeneous preparation.

 

Rinsing Solution Preparation:

In a suitable beaker, transferred ~500mL purified water with 20mL concentrated HNO₃ and 10mL concentrated HCl. Added 0.3mL gold standard (1000µg/mL), then diluted to 1000mL with water and mixed uniformly for homogeneity.

Note: Use micropipette for all linearity solution preparation

 

Internal Standard Preparation:

Internal Standard Stock Solution:

Added 1.6mL Yttrium standard, 0.45mL Indium standard, and 0.16mL Bismuth standard to a 100mL volumetric flask. Brought to volume using Diluent-1 and mixed thoroughly.

Internal Standard Solution:

Pipetted 1.5mL Germanium standard (1000µg/mL) and 5.0mL internal standard stock into a 100mL volumetric flask. Diluted to mark with Diluent-1 and mixed uniformly.

 

Linearity Solution Preparation:

Linearity Stock Solution-1:

Added 0.926mL Vanadium standard (1000µg/mL), 0.463mL Cobalt standard (1000µg/mL), and 1.852mL Nickel standard (1000µg/mL) to a 50mL volumetric flask. Filled to volume with Diluent-1 and mixed thoroughly.

Linearity Stock Solution-2:

Pipetted 1.390mL Arsenic standard (1000 µg/mL), 0.463 mL Cadmium standard (1000 µg/mL), 2.780 mL Mercury standard (1000µg/mL), and 0.463mL Lead standard (1000µg/mL) into a 100mL volumetric flask. Brought to volume with Diluent-1 and mixed well.

 

Linearity Stock Solution-3:

Transferred 5.0 mL Linearity stock solution-1 and 1.0 mL Linearity stock solution-2 in a 50 mL volumetric flask. Diluted to mark with Diluent-1 and mixed thoroughly.


 

Linearity Solution Preparation:

Table 4: Preparation scheme for linearity calibration solutions used in ICP–MS analysis.

Level

Linearity stock Solution-3 to be taken (mL)

Internal Standard Solution to be taken (mL)

Dilute to volume with Diluent-1 (mL)

Concentration in ppb

V

Co

Ni

As

Cd

Hg

Pb

Solution A (10 %)

0.1

1.0

50

3.704

1.852

7.408

0.556

0.185

1.112

0.185

Solution B

(25 %)

0.25

1.0

50

9.260

4.630

18.520

1.390

0.463

2.780

0.463

Solution C

(50 %)

0.5

1.0

50

18.520

9.260

37.040

2.780

0.926

5.560

0.926

Solution D

(100 %)

1.0

1.0

50

37.040

18.520

74.080

5.560

1.852

11.120

1.852

Solution E

(150 %)

1.5

1.0

50

55.560

27.780

111.120

8.340

2.778

16.680

2.778

Solution F

(200 %)

2.0

1.0

50

74.080

37.040

148.160

11.120

3.704

22.240

3.704

Bias and Drift

Use Solution E (150 %) for bias and drift check

 

 


Blank solution:

Transferred 1.0mL Internal Standard Solution to a 50 mL volumetric flask, diluted to volume with Diluent-1, and mixed well.

Calibration blank

Blank solution served as calibration blank.

 

Sample Blank Solution: Sample blank stock was made by adding 10.0mL concentrated nitric acid in a digestion vessel, left open at ambient temperature for 10minutes with occasional swirling. Next, added 0.1mL concentrated hydrofluoric acid, then held open again for 10minutes with intermittent mixing. Sealed vessels underwent microwave digestion following programmed parameters. After digestion, transferred contents quantitatively to a 100mL volumetric flask. Rinsed vessel three times (~15 mL Diluent-2 each), combining rinsates in the flask. Diluted to volume with Diluent-2, mixed thoroughly, filtered via 0.45µm PVDF (discarding first 5mL), and stored for analysis. To prepare the final blank, transferred 5mL stock solution to a 25mL volumetric flask with 0.5mL internal standard. Filled to mark using Diluent-1, mixed uniformly, and readied for ICP-MS24.

 

Sample Preparation: Tablets were crushed finely with a grinder and blended evenly for uniformity. Precisely weighed ~0.2g powdered sample into a digestion vessel, added 10.0mL concentrated nitric acid, and let stand open at room temperature for 10min with periodic swirling. Then introduced 0.1mL concentrated hydrofluoric acid, holding open for another 10min with intermittent mixing. Sealed the vessel and processed via microwave digestion using defined program settings. Post-digestion, vessels were cooled, then contents quantitatively transferred to a 100mL volumetric flask. Rinsed the vessel thrice with ~15mL Diluent-2 portions, adding all rinsings to the flask. Diluted to volume with Diluent-2, mixed thoroughly, then passed through a 0.45µm PVDF filter, discarding initial 5mL filtrate. For the final sample blank, pipetted 5mL stock solution into a 25mL volumetric flask, added 0.5mL internal standard, diluted to volume with Diluent-1, and mixed well before analysis.

 

The aspiration sequence consisted of the blank solution, calibration blank, and linearity solutions A through F, followed by the sample preparations. Blank solutions were aspirated at regular intervals throughout the sequence to effectively eliminate any potential carryover and ensure measurement reliability24.

Method Validation:

Validation covered specificity, LOD/LOQ, linearity, precision (system and method), accuracy, and system suitability, meeting USP <233>, <730> and ICH guidelines14,10. Table 10 lists targeted limits (μg/g) derived from PDE values (μg/day) for Class 1, 2A, and 2B elemental impurities25,26.


 

Table 5: Oral PDE limits and corresponding concentration limits (µg/g) for selected elemental impurities

Name of Element

Class

Oral PDE μg/Day spiking sample solutions with known elemental impurities at LOQ to 150% levels

μg/g (as per 10g/day)

Vanadium (V)

2A

100

10

Cobalt (Co)

2A

50

5

Nickel (Ni)

2A

200

20

Arsenic (As)

1

15

1.5

Cadmium (Cd)

1

5

0.5

Mercury (Hg)

1

30

3

Lead (Pb)

1

5

0.5

 


Specificity:

Method specificity was evaluated by comparing the signal responses obtained from blank preparations with those obtained from spiked sample solutions containing elemental impurities at the LOQ level. For all investigated elements, the background response observed in the blank solution was significantly lower than the corresponding signal generated from the spiked samples. This demonstrated the absence of significant spectral or matrix interferences at the selected analyte masses. The large difference between blank and spiked responses confirmed that the analytical method was capable of selectively detecting and quantifying each elemental impurity in the presence of the sample matrix without interference from reagents, excipients, or instrumental background signals.

 

Linearity:

The linearity of vanadium, cobalt, nickel, arsenic, cadmium, mercury, and lead was established by analysing calibration standards prepared at six concentration levels corresponding to 10%-200% of the working standard concentration. The evaluated concentration ranges were 3.715–74.302 ppb for vanadium, 1.852–37.040 ppb for cobalt, 7.408–148.160 ppb for nickel, 0.554–11.086 ppb for arsenic, 0.185–3.700 ppb for cadmium, 1.116–22.328 ppb for mercury, and 0.185–3.700 ppb for lead. Each calibration solution was analysed using ICP-MS, and calibration plots were generated by correlating analyte concentration with the corresponding detector response (CPS). The obtained results indicated consistent detector response and satisfactory linear behaviour throughout the studied range for all investigated elemental impurities. Calibration curves for V, Co, Ni, As, Cd, Hg, and Pb demonstrated strong proportionality between concentration and response, with correlation coefficient (R) values ≥0.9977 and R˛ values ≥0.9955, confirming reliable and linear detector performance across the studied range.

 

Figure 1. Instrument Derived Linearity plot

 

All target elements showed correlation coefficients (R) ≥0.990, confirming strong linearity. The method remained linear across 10-200% of the working concentration for V, Co, Ni, As, Cd, Hg, and Pb. These results confirm a strong and proportional relationship between analyte concentration and instrumental response across the studied range.

 

LOD Results:

The sensitivity of the developed ICP-MS method was assessed through determination of the limits of detection (LOD) for vanadium, cobalt, nickel, arsenic, cadmium, mercury, and lead. The obtained LOD values were 2.972 ppb for vanadium, 1.481 ppb for cobalt, 5.926 ppb for nickel, 0.443ppb for arsenic, 0.148ppb for cadmium, 0.893ppb for mercury, and 0.148ppb for lead. In terms of concentration in the sample matrix, the corresponding values were found to be 0.002972, 0.001481, 0.005926, 0.000443, 0.000148, 0.000893, and 0.000148μg/g, respectively. The calculated daily exposure levels were within acceptable limits for all analysed elemental impurities. Blank solution responses showed minimal background interference, while the analyte responses at the LOD level produced distinguishable and measurable signal intensities, confirming adequate method sensitivity. These findings demonstrate that the proposed ICP-MS method is capable of detecting trace levels of elemental impurities with satisfactory precision and reliability, making it suitable for routine low-level elemental analysis in pharmaceutical samples.

 

LOQ Results:

The limits of quantification (LOQ) for vanadium, cobalt, nickel, arsenic, cadmium, mercury, and lead were established to evaluate the quantitative capability of the developed ICP-MS method at low concentration levels. The obtained LOQ values were 9.287ppb for vanadium, 4.630 ppb for cobalt, 18.520 ppb for nickel, 1.385ppb for arsenic, 0.462ppb for cadmium, 2.791 ppb for mercury, and 0.462ppb for lead. The corresponding concentrations in the sample matrix were calculated as 0.009287, 0.004630, 0.018520, 0.001385, 0.000462, 0.002791, and 0.000462μg/g, respectively. Detector responses at the LOQ level were significantly higher than those obtained for blank solutions, indicating clear signal discrimination and acceptable analytical sensitivity. Precision at the LOQ level was assessed using six replicate measurements, and the %RSD values for all elemental impurities were found to be below 10%, demonstrating satisfactory repeatability of the method. The results confirm that the developed ICP-MS procedure is suitable for accurate and reliable quantification of trace elemental impurities at low concentration levels during routine pharmaceutical analysis.

 

Precision:

System Precision:

System precision (standard linearity) of the ICP-MS method demonstrated excellent performance for all evaluated elements, with correlation coefficients ranging from 0.9961 to 0.9999. These values confirm a strong linear relationship between concentration and instrumental response. Drift assessment at 150% level revealed %RSD of 0.4-2.7% from Check-1 to Check-2, evidencing signal constancy throughout the sequence. System precision drift was evaluated using Solution E (150%) during the analytical sequence. The observed drift for all elemental impurities ranged from 0.4% to 2.7%, indicating stable instrument performance and consistent signal response throughout the ICP-MS analysis.

 

Method Precision:

Six replicate samples, each spiked with target elemental impurities at specification levels, were prepared and tested per the procedure. Results appear in Table 6.

 

Accuracy:

Method accuracy was verified by spiking sample solutions with known elemental impurities at LOQ to 150% of specification levels for V, Co, Ni, As, Cd, Hg, and Pb, then analyzing per the procedure. Results for these elements are in Table 7.


 

 

 

Table 6. ICP-MS method precision for elemental analysis (%RSD, mean, min and max)

Sample

Vanadium (V)

Cobalt (Co)

Nickel (Ni)

Arsenic (As)

Cadmium d)

Mercury (Hg)

Lead (Pb)

Mean

77.060

37.766

151.517

12.444

3.769

22.690

3.952

Minimum

77.076

37.523

150.960

12.333

3.763

22.424

3.894

maximum

76.899

37.897

152.078

12.626

3.776

22.907

4.025

%RSD

0.4

0.3

0.2

0.9

0.1

0.7

1.5

USP precision (repeatability) criteria were satisfied, with %RSD ≤20% across six spiked preparations per element. This confirms the procedure's reliability for method precision and suitability for intended application.

 


Table 7. ICP-MS accuracy (% recovery and %RSD) at LOQ, 50%, 100%, and 150% levels

Name of Element

LOQ%

50%

100%

150%

Overall mean % recovery

Overall %RSD

Mean Recovery (%)

%RSD

Mean Recovery (%)

%RSD

Mean Recovery (%)

%RSD

Mean Recovery (%)

%RSD

Vanadium (V)

91.7

0.5

90.8

0.8

91.4

0.1

93.9

1.8

92.0

1.6

Cobalt (Co)

89.6

0.4

88.8

0.2

90.0

0.1

91.6

3.7

90.0

2.0

Nickel (Ni)

90.1

0.6

90.3

0.4

90.9

0.5

91.8

1.1

90.8

0.9

Arsenic (As)

99.0

0.6

97.8

0.4

97.7

0.7

97.2

1.0

97.8

0.9

Cadmium (Cd)

98.3

0.9

98.8

1.1

97.8

0.9

98.2

1.3

98.3

1.0

Mercury (Hg)

100.6

1.2

98.4

0.3

98.8

0.3

97.9

0.8

98.9

1.2

Lead (Pb)

100.4

1.9

99.9

2.5

98.9

0.6

97.6

0.5

99.2

1.8

Mean recoveries per level and overall, for each target element fell within 70.0–150.0%. This fulfilled USP <233/232> accuracy requirements.

 


ROBUSTNESS RESULTS:

Robustness of the ICP-MS method was evaluated by varying the acid volume (±1mL) from the nominal condition during sample preparation. The impact was assessed by comparing measured concentrations and percentage differences for all elements. Results showed minimal variation in analyte response under altered conditions. % Differences stayed below 10% (NMT limit), showing no impact on results. Signals avoided bias no suppression or boost. Method holds up well for regular impurity checks.


 

Table 8. ICP-MS robustness results showing % difference with variation in acid volume

Element

ppb of 10mL Acid volume (As per method)

9 mL acid volume

11 mL acid volume

ppb

% Difference (NMT 10%)

ppb

% Difference (NMT 10%)

Vanadium (V)

33.336

33.212

0.4

33.418

0.2

Cobalt (Co)

16.668

16.417

1.5

16.702

0.2

Nickel (Ni)

66.672

66.528

0.2

66.660

0.0

Arsenic (As)

5.393

5.320

1.4

5.408

0.3

Cadmium (Cd)

1.814

1.791

1.3

1.821

0.4

Mercury (Hg)

11.008

10.992

0.1

10.989

0.2

Lead (Pb)

1.833

1.802

1.7

1.852

1.0

 

 


CONCLUSION:

We developed and validated an ICP-MS method sensitive, precise and rugged for Deflazacort tablet impurities. It matches USP <232/233> and ICH Q3D rules. Dilute nitric acid microwave digestion gave clean, consistent preparations with low contamination. Validation passed specificity through LOD, plus accuracy and precision checks. Measured levels stayed under oral PDE limits, supporting everyday QC use.

 

FUTURE RECOMMENDATIONS:

Researchers could apply this ICP-MS approach to additional Deflazacort formats and related steroids. Ongoing ICH Q3D risk evaluations would track shifts in suppliers or production. Work on better interference mitigation via collision/reaction cells might improve performance. Stability tests over time could check leaching from containers. Multi-site validation would aid method adoption worldwide.

 

PRACTICAL IMPLICATIONS:

QC labs gain a sensitive multi-element ICP-MS tool from this validation for meeting USP/ICH standards. Routine testing gets faster and cheaper, with safer microwave prep using less acid. It bolsters impurity monitoring across production stages to protect end users.

 

REFERENCES:

1.      Analytical approach of elemental impurities in pharmaceutical products: A worldwide review. DOI: 10.1016/j.sab.2023.106689

2.      Range and Output Simulation for Elemental Impurities in Drug Products

3.      ICH Q3D Drug Product Elemental Risk Assessment: The Use of An Elemental Impurities Excipients Database. DOI: org/10.1016/ j.xphs.2021.10.012

4.      The Development of Analytical Procedures for Analysis of Trace Metals in Pharmaceutical Formulations and the Speciation of Arsenic in Antacids.

5.      ICH-Q3D based elemental impurities study in an oral drug product: Risk assessment according to the component approach (option 2b) confirmed by the finished product analysis (option 3) DOI: 10.1016/ j.pharma.2025.05.002

6.      Method Development and Validation for Detection of Selective Elemental Impurities in Clarithromycin API by ICP-MS, DOI:10.36348/sijtcm.2026.v09i01.001s

7.      The comparative verification of calibration curve and background fundamental parameter methods for impurity analysis in drug materials. DOI: org/10.1002/ xrs.2788

8.      Quantitative Determination of Essential and Trace Element Content of Some Medicinal Plants by ICP-MS Technique. DOI: 10.5958/0974-360X.2019.00265.8

9.      Method Development and Validation for Simultaneous Determination of 44Ca, 34S, 28Si and 18 other Trace Elements

<233> elemental impurities-procedures

10.   Elemental Impurities Estimation in Adenosine Injection using ICP-MS. DOI: 10.52711/0974-360X.2025.00189

11.   Simultaneous Estimation of Fenofibric Acid Tablets' Elemental Impurities using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) DOI:10.52711/2231-5675.2025.00001

12.   Flow injection combined with ICP-MS for accurate high throughput analysis of elemental impurities in pharmaceutical products according to USP <232>/<233>. DOI: 10.1016/j.jpba.2014.02.016

13.   Guideline for elemental impurities Q3D(R2)

14.   Development and validation of elemental impurities by the ICP-MS method in phenoxybenzamine hydrochloride capsules. DOI: DOI: org/10.24327/ijrsr.2021.1201.5694

15.   Optimisation of operating parameters for inductively coupled plasma mass spectrometry: A computational study, DOI: org/10.1016/ j.sab.2012.06.006

16.   ICP-MS: a tool for detection and quantitation of fosfomycin residues in cleaning samples of finished product by estimation of phosphorous load, DOI: 10.1039/D3JA00247K

17.   Determination of Elemental Impurities in Etoricoxib a Non-steroidal Anti-inflammatory Drug by using ICP-MS. DOI: 10.52711/0974-360X.2026.00155

18.   Microwave-assisted sample preparation of medicines for the determination of elemental impurities in compliance with the United States Pharmacopoeia. DOI: org/10.1016/j.aca.2019.03.016

19.   Elemental impurities analysis in name-brand and generic omeprazole drug samples. DOI: org/10.1016/j.heliyon. 2020.e03359

20.   Internal standards in inductively coupled plasma mass spectrometry using kinetic energy discrimination and dynamic reaction cells, DOI: 10.1039/ C8JA00171E

21.   Enhanced ICP-MS detection of 24 elemental impurities in complex, high matrix mineral, medicinal lanthanum carbonates according to ICH Q2 (R2). DOI: org/10.1039/D4AY01589D

22.   Determination of elemental impurities in pharmaceutical products and related matrices by ICP-based methods: a review. DOI 10.1007/ s00216-016-9471-6

23.   Analytical validation of a microwave-assisted ICP-MS method for the quantification of elemental impurities in doxycycline Hyclate tablets. DOI: org/10.1007/s11696-024-03467-6

24.   Determination of elemental impurities of Arsenic, Cadmium, Mercury, Lead and Palladium content in Testosterone propionate by using ICP-MS. DOI: 10.52711/0974-4150.2021.00035

25.   Determination of ICH class 1 and class 2A trace elements in Triheptanoin oral liquid for long-chain fatty acid oxidation disorders, an ICP-MS approach, DOI:10.1007/s44371-025-00302-z

 

 

 

Received on 14.05.2026      Revised on 11.06.2026

Accepted on 03.07.2026      Published on 12.10.2026

Available online from October 21, 2026

Asian Journal of Pharmaceutical Analysis. 2026; 16(4):265-273.

DOI: 10.52711/2231-5675.2026.00041

©Asian Pharma Press All Right Reserved

 

This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License. Creative Commons License.